Dynamic Load Compensation (DLC) System

Brand: JOFRA Calibration

The DLC Dynamic Load Compensation Systems is yet another state of the art innovation within temperature calibration.

The DLC compensates for the negative effect on calibration accuracy that heat exchange between a dry-block calibrator and the sensor-under-test creates.

When a sensor is inserted into a dry-block calibrator, the sensor size (diameter and length) and the difference between calibration temperature and ambient temperature, create an axial gradient deviation inside the insert.

Our temperature calibrators are already famous for their active dual-zone calibration principle, and the DLC takes this one step further, and now compensates inside the insert.

The DLC, combined with an RTC Calibrator, creates the only dry-block that compensates for the heat dissipation caused by sensor loading.
Features
  • Control the lower section of the insert to within a few hundredths of a degree
  • The DLC indicator shows that the dual-zone is active and working
  • A DLC checkmark is present when temperature homogeneity is achieved
  • Save time by calibrating several sensors simultaneously
  • Calibrate large diameter sensors without losing accuracy
  • Aligning the thermos-sensitive part of the sensor is no longer needed

Specifications

Probe DLC-155

  • Sensor Type: Differential TC Type T
  • Probe Length: 177 mm (6.96 in)
  • Temperature Range: -60° C to 155° C / -76° F to 311° F
  • Diameter: O.D. 3 mm
  • Depth: 140 mm (5.51 in)
  • Media Compatibility: AISI316
  • Connection: Redel Plug with Built in Memory is Standard

Probe DLC-158

  • Sensor Type: Differential TC Type T
  • Probe Length: 196 mm (7.72 in)
  • Temperature Range: -60° C to 155° C / -76° F to 311° F
  • Diameter: O.D. 3 mm
  • Depth: 140 mm (5.51 in)
  • Media Compatibility: AISI316
  • Connection: Redel Plug with Built in Memory is Standard

Probe DLC-159

  • Sensor Type: Differential TC Type T
  • Probe Length: 196 mm (7.72 in)
  • Temperature Range: -100° C to 155° C / -148° F to 311° F
  • Diameter: O.D. 3 mm
  • Depth: 140 mm (5.51 in)
  • Media Compatibility: AISI316
  • Connection: Redel Plug with Built in Memory is Standard

Probe DLC-168

  • Sensor Type: Differential TC Type T
  • Probe Length: 196 mm (7.72 in)
  • Temperature Range: -60° C to 165° C / -76° F to 329° F
  • Diameter: O.D. 3 mm
  • Depth: 140 mm (5.51 in)
  • Media Compatibility: AISI316
  • Connection: Redel Plug with Built in Memory is Standard

Probe DLC-187

  • Sensor Type: Differential TC Type T
  • Probe Length: 177 mm (6.96 in)
  • Temperature Range: -60° C to 180° C / -76° F to 356° F
  • Diameter: O.D. 3 mm
  • Depth: 140 mm (5.51 in)
  • Media Compatibility: AISI316
  • Connection: Redel Plug with Built in Memory is Standard

Probe DLC-250

  • Sensor Type: Differential TC Type T
  • Probe Length: 196 mm (7.72 in)
  • Temperature Range: 0° C to 250° C / 32° F to 482° F
  • Diameter: O.D. 3 mm
  • Depth: 140 mm (5.51 in)
  • Media Compatibility: AISI316
  • Connection: Redel Plug with Built in Memory is Standard

Probe DLC-700

  • Sensor Type: Differential TC Type N
  • Probe Length: 230 mm (9.05 in)
  • Temperature Range: 0° C to 700° C / 32° F to 1292° F
  • Diameter: O.D. 4 mm
  • Depth: 200 mm (7.9 in)
  • Media Compatibility: INCONEL 600
  • Connection: Redel Plug with Built in Memory is Standard

  • Ultrasonic Flow Meters for Custody Transfer of Natural Gas
    • The Ability to Accurately Measure the Flow of Gas in the Pipeline is Critical for Custody Transfer

Show More